Please use this identifier to cite or link to this item:
https://idr.l1.nitk.ac.in/jspui/handle/123456789/14717
Title: | Design of bypass rotary vane magneto rheological damper for prosthetic knee application |
Authors: | Tak R.S.S. Kumar H. Chandramohan S. Srinivasan S. |
Issue Date: | 2019 |
Citation: | Proceedings of 30th International Conference on Adaptive Structures and Technologies, ICAST 2019 , Vol. , , p. 105 - 106 |
Abstract: | In this paper, a bypass rotary vane type magneto rheological (MR) damper suitable for prosthetic knee application is designed. The torque and angular velocity requirements of the normal human knee are used as design limits. In the proposed design, the rotary vane chamber and the MR valve are connected by hydraulic cables and ports and are designed separately. The rotary vane chamber is designed based on the cross-sectional size limits of the normal human knee, while the MR valve is designed with the objective of obtaining the maximum on-state damping torque, minimum weight, and minimum off-state torque. © copyright Environment and Climate Change Canada. |
URI: | https://doi.org/ http://idr.nitk.ac.in/jspui/handle/123456789/14717 |
Appears in Collections: | 2. Conference Papers |
Files in This Item:
There are no files associated with this item.
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.